Fluorescence Chemical Transducers




The knowledge of the partial pressure of oxygen (PO2), the pH, and the partial pressure of carbon dioxide (PCO2) in arterial blood is very important in the management of critically ill patients. Sensing elements with specific reagent chemistries for each of the above measurands are attached to the tips of each of three optical fibers. When used as intravascular transducers, the three fibers are enclosed in a single catheter which is then inserted into the artery through an incision for real-time continuous monitoring of the blood gases and pH. Light from a laser diode source at a specific wavelength enters the fiber at one end (proximal), propagates through the core, and irradiates the sensing element at the distal end inside the artery.

The sensing element contains a dye with fluorescent molecules which, on absorption of the photons, are excited to a higher-energy state. In that state, the molecule loses some of its energy to the lattice structure. As it returns with lower energy from an intermediate state of excitation, it fluoresces at a different wavelength.

 Since the energy of a photon is inversely proportional to the wavelength, the energy in the emitted light is of a longer wavelength. The presence of an analyte (measurand) modifies the intensity of the reemitted (fluorescent) light. It can be measured using wavelength-selective devices and photonic detectors, whose  output is proportional to the concentration of the measurand. An analyte such as oxygen quenches the fluorescence intensity.

This is the basic transduction mechanism in PO2 transducers. Based on the fluorescence quenching effect, intravascular PO2 transducers have been built as shown in Fig. 5.47a. The relationship between PO2 (measured) and the fluorescence characteristics of the dye molecule is given by the rearranged form of the Stern-Voltmer equation



where I is the fluorescence intensity after quenching by oxygen and I0 is the intensity in the absence of oxygen. K is the quenching constant, and it represents the slope of the PO2 vs. I0/I plot, which characterizes the performance of the oxygen transducer.108 By measuring I0 at the excitation wavelength and at the fluorescence wavelength , the partial pressure of oxygen can be determined.
The sensing element in a pH transducer is a pH-sensitive dye, whose acid from (HA) and the base form (A_) fluoresce at different excitation wavelengths, but the longer wavelength of fluorescence emission is the same for both forms,



Figure 5.47 Fiber-optic fluorescence chemical transducers. (a) Oxygen transducer based on the fluorescence quenching effect. (b) pH transducer.
as indicated in Fig. 5.47b.
This is the basic transduction mechanism. In this case, the pH can be expressed as


Physical Transducers




The most direct application of an optical fiber to measure physical measurands is the microbend transducer, as shown in Fig. 5.44.102,103,105 fiber is deformed by a measurand such as displacement, strain, pressure, force, acceleration, or temperature. In each case, the transduction mechanism is the decrease in light intensity of the beam through the fiber core as a result of the deformation, which causes the core modes to lose energy to the cladding modes. A photodetector is used to provide a corresponding electrical output. This is an intrinsic-type transducer.

Figure 5.45 shows a pressure transducer and a temperature transducer


Fig 5.45 Fiber-optic temperature and pressure transducers.


configured as a Fabry-Perot etalon or interferometer. The spacing between mirrors is critical to the operation of the etalon as a physical transducer.

Any change in that spacing, brought about by a measurand, will cause the transmittance of the etalon to change in proportion to the measurand. These extrinsic transducers are small in size and can be constructed by using integrated circuit techniques.

Optical fibers are also used to measure magnetic and acoustic fields. In a magnetic field transducer as shown in Fig. 5.46, the transduction mechanism consists of the longitudinal strain in the magnetostrictive material caused by the magnetic field and a corresponding strain in the core.

This strain modulates the refractive index of the core and produces a field-dependent phase shift in the output optical beam. An acoustic field transducer can also be made along these lines by replacing the magnetostrictive substrate with an acoustically sensitive coating around the fiber cladding. Consequently, this produces an acoustic field-dependent phase shift of the optical beam.


Fig 5.46 Fiber-optic magnetic field transducer.

A Mach-Zhender interferometer is used to obtain an electrical output proportional to the measurands.

Position Sensing Transducers




Structurally modified photodiodes are used in the measurement of small angles, distances, and machine vibrations, which are related to linear position sensing.

Fig 5.40 Position sensing transducer.


The lateral effect PIN photodiode is generally used for these measurements, and its cross section is shown in Fig. 5.40a. Photons from a laser source impinge on the p-type front surface of the PIN diode and are readily absorbed within the depletion region, where electron-hole pairs are generated.

The electrons move toward the n-type region and generate a photocurrent. The operation of the lateral-effect photodiode can be represented by a variable contact resistor, as shown in Fig. 5.40b. The location of the laser spot defines that variable contact. The photocurrent I0 divides between the paths to the two edge electrodes in a ratio given by



where L is the distance between the electrodes, S is the distance of the laser spot from one of the electrodes, and I0=I1+I2. Using the above equation, S can be determined by measuring I1 and I2. This is the transduction mechanism of the position sensing transducer. It measures displacements along a single axis.

Physical Transducers.





In transducers used for force, acceleration, and temperature measurements, the length L is modified by the measurands. In crystalline quartz devices, the fractional change in SAW velocity is small compared with the surface strain caused by a change in L.

We can therefore expect a decrease in oscillator frequency when L is increased by direct axial tension or by an increase in temperature.So, in transducers that measure force, the temperature must be maintained constant. Figure 5.27 shows a dual oscillator configuration where the temperature effects are reduced and the force sensitivity is doubled. For a force acting downward, the A1 oscillator increases its frequency and the A2 oscillator decreases its frequency.

Figure 5.27 Dual SAW-oscillator force transducer. (a) Oscillators on opposite faces of quartz substrate achieve temperature compensation. (b) Change in oscillator frequency as a function of the force. Mixer output shows almost double the sensitivity.

The difference in frequency is a measure of the force.


Labels

PROJECTS 8086 PIN CONFIGURATION 80X86 PROCESSORS TRANSDUCERS 8086 – ARCHITECTURE Hall-Effect Transducers INTEL 8085 OPTICAL MATERIALS BIPOLAR TRANSISTORS INTEL 8255 Optoelectronic Devices Thermistors thevenin's theorem MAXIMUM MODE CONFIGURATION OF 8086 SYSTEM ASSEMBLY LANGUAGE PROGRAMME OF 80X86 PROCESSORS POWER PLANT ENGINEERING PRIME MOVERS 8279 with 8085 MINIMUM MODE CONFIGURATION OF 8086 SYSTEM MISCELLANEOUS DEVICES MODERN ENGINEERING MATERIALS 8085 Processor- Q and A-1 BASIC CONCEPTS OF FLUID MECHANICS OSCILLATORS 8085 Processor- Q and A-2 Features of 8086 PUMPS AND TURBINES 8031/8051 MICROCONTROLLER Chemfet Transducers DIODES FIRST LAW OF THERMODYNAMICS METHOD OF STATEMENTS 8279 with 8086 HIGH VOLTAGE ENGINEERING OVERVOLATGES AND INSULATION COORDINATION Thermocouples 8251A to 8086 ARCHITECTURE OF 8031/8051 Angle-Beam Transducers DATA TRANSFER INSTRUCTIONS IN 8051/8031 INSTRUCTION SET FOR 8051/8031 INTEL 8279 KEYBOARD AND DISPLAY INTERFACES USING 8279 LOGICAL INSTRUCTIONS FOR 8051/8031 Photonic Transducers TECHNOLOGICAL TIPS THREE POINT STARTER 8257 with 8085 ARITHMETIC INSTRUCTIONS IN 8051/8031 LIGHTNING PHENOMENA Photoelectric Detectors Physical Strain Gage Transducers 8259 PROCESSOR APPLICATIONS OF HALL EFFECT BRANCHING INSTRUCTIONS FOR 8051/8031 CPU OF 8031/8051 Capacitive Transducers DECODER Electromagnetic Transducer Hall voltage INTEL 8051 MICROCONTROLLER INTEL 8251A Insulation Resistance Test PINS AND SIGNALS OF 8031/8051 Physical Transducers Resistive Transducer STARTERS Thermocouple Vacuum Gages USART-INTEL 8251A APPLICATIONs OF 8085 MICROPROCESSOR CAPACITANCE Data Transfer Instructions In 8086 Processors EARTH FAULT RELAY ELECTRIC MOTORS ELECTRICAL AND ELECTRONIC INSTRUMENTS ELECTRICAL BREAKDOWN IN GASES FIELD EFFECT TRANSISTOR (FET) INTEL 8257 IONIZATION AND DECAY PROCESSES Inductive Transducers Microprocessor and Microcontroller OVER CURRENT RELAY OVER CURRENT RELAY TESTING METHODS PhotoConductive Detectors PhotoVoltaic Detectors Registers Of 8051/8031 Microcontroller Testing Methods ADC INTERFACE AMPLIFIERS APPLICATIONS OF 8259 EARTH ELECTRODE RESISTANCE MEASUREMENT TESTING METHODS EARTH FAULT RELAY TESTING METHODS Electricity Ferrodynamic Wattmeter Fiber-Optic Transducers IC TESTER IC TESTER part-2 INTERRUPTS Intravascular imaging transducer LIGHTNING ARRESTERS MEASUREMENT SYSTEM Mechanical imaging transducers Mesh Current-2 Millman's Theorem NEGATIVE FEEDBACK Norton's Polarity Test Potentiometric transducers Ratio Test SERIAL DATA COMMUNICATION SFR OF 8051/8031 SOLIDS AND LIQUIDS Speed Control System 8085 Stepper Motor Control System Winding Resistance Test 20 MVA 6-digits 6-digits 7-segment LEDs 7-segment A-to-D A/D ADC ADVANTAGES OF CORONA ALTERNATOR BY POTIER & ASA METHOD ANALOG TO DIGITAL CONVERTER AUXILIARY TRANSFORMER AUXILIARY TRANSFORMER TESTING AUXILIARY TRANSFORMER TESTING METHODS Analog Devices A–D BERNOULLI’S PRINCIPLE BUS BAR BUS BAR TESTING Basic measuring circuits Bernoulli's Equation Bit Manipulation Instruction Buchholz relay test CORONA POWER LOSS CURRENT TRANSFORMER CURRENT TRANSFORMER TESTING Contact resistance test Current to voltage converter DAC INTERFACE DESCRIBE MULTIPLY-EXCITED Digital Storage Oscilloscope Display Driver Circuit E PROMER ELPLUS NT-111 EPROM AND STATIC RAM EXCITED MAGNETIC FIELD Electrical Machines II- Exp NO.1 Energy Meters FACTORS AFFECTING CORONA FLIP FLOPS Fluid Dynamics and Bernoulli's Equation Fluorescence Chemical Transducers Foil Strain Gages HALL EFFECT HIGH VOLTAGE ENGG HV test HYSTERESIS MOTOR Hall co-efficient Hall voltage and Hall Co-efficient High Voltage Insulator Coating Hot-wire anemometer How to Read a Capacitor? IC TESTER part-1 INSTRUMENT TRANSFORMERS Importance of Hall Effect Insulation resistance check Insulator Coating Knee point Test LEDs LEDs Display Driver LEDs Display Driver Circuit LM35 LOGIC CONTROLLER LPT LPT PORT LPT PORT EXPANDER LPT PORT LPT PORT EXTENDER Life Gone? MAGNETIC FIELD MAGNETIC FIELD SYSTEMS METHOD OF STATEMENT FOR TRANSFORMER STABILITY TEST METHODS OF REDUCING CORONA EFFECT MULTIPLY-EXCITED MULTIPLY-EXCITED MAGNETIC FIELD SYSTEMS Mesh Current Mesh Current-1 Moving Iron Instruments Multiplexing Network Theorems Node Voltage Method On-No Load And On Load Condition PLC PORT EXTENDER POTIER & ASA METHOD POWER TRANSFORMER POWER TRANSFORMER TESTING POWER TRANSFORMER TESTING METHODS PROGRAMMABLE LOGIC PROGRAMMABLE LOGIC CONTROLLER Parallel Port EXPANDER Paschen's law Piezoelectric Wave-Propagation Transducers Potential Transformer RADIO INTERFERENCE RECTIFIERS REGULATION OF ALTERNATOR REGULATION OF THREE PHASE ALTERNATOR Read a Capacitor SINGLY-EXCITED SOLIDS AND LIQUIDS Classical gas laws Secondary effects Semiconductor strain gages Speaker Driver Strain Gages Streamer theory Superposition Superposition theorem Swinburne’s Test TMOD TRANSFORMER TESTING METHODS Tape Recorder Three-Phase Wattmeter Transformer Tap Changer Transformer Testing Vector group test Virus Activity Voltage Insulator Coating Voltage To Frequency Converter Voltage to current converter What is analog-to-digital conversion Windows work for Nokia capacitor labels excitation current test magnetic balance voltage to frequency converter wiki electronic frequency converter testing voltage with a multimeter 50 hz voltages voltmeter